Chapter 6
|
|
|
FIGURE 6-25 This high resolution image of a test grid shows the effect of acoustic noise on an image. A: Image and line profiles measured while acoustic noise was present in the room. B: Image that was measured without the acoustic noise. |
6.5 Other Sources |
6.5.1 Surface Contamination |
Substantial contamination at the surface of a sample such as a fi ngerprint or oil film can cause AFM image artifacts. Such artifacts appear as streaks on the image especially in locations where there are "sharp" features and edges on the sample's surface. Often the streaking can be reduced or even
eliminated by cleaning the sample with a high purity solvent. |
|
FIGURE 6-26 A: SEM image of a test pattern that is contaminated. B: AFM image of the same test pattern that is covered with contamination. The contamination is identified by the streak marks at the top of the scan. |
6.5.2 Electronics |
Image artifacts can appear in AFM scans because of faulty electronics. |
|
|